Skip to Main Content
Site IDParent SiteLocationCountyEFO
4808AOC, LLC dba AOC Tennessee Gelcoats and Colorants Plant860 SR- 57 EastFayetteMemphis

Inspection Details

67995
4808
TN0000442
2/26/2020
Compliance Eval (non-sampling) (CEI)
Out of Compliance
During the inspection the DO meter was unable to read the proper saturation in the air saturated water that is used for calibration. The meter should have read near 100%, but never registered that value after several attempts.

Calibration records of the pH and DO meters are not maintained.

It was observed that the BOD Nutrient Buffer Pillows that are used for BOD analysis had expired in April 2017 and the TRC pillows had expired in November 2016. Using expired buffers is not a practice that is recognized in the Standard Method.